Directed Electronics 1500X User's Guide Page 7

  • Download
  • Add to my manuals
  • Print
  • Page
    / 30
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 6
AngioDynamics, Inc. Model 1500X User’s Guide and Service Manual
160-103990 Rev 03 Page 5
Cables connected to the RITA
Device should not contact the patient or other
electrical leads.
Skin-to-skin contact, such as between the torso and the arms, or between the legs of
the patient should be avoided by insulating these contacts with sheets or dry towels.
Failure of high frequency surgical equipment could result in an unintended increase
of output power.
When not in use electrosurgical leads (active or return) should be positioned so that
they cannot come into contact with the patient or other leads.
High power settings can cause local desiccation of tissue, which can impede the
ability to produce expected ablations. Set power as low as possible for intended
purpose. Follow manufacturer’s guidelines of time at temperature for ablation
generation. If the recommended times and temperatures are not achieved at full
deployment of the Electrosurgical Device array, there can be no assurance that the
desired ablation volume has been created. Standard evaluative techniques, e.g., CT
or MRI, should be used to determine the actual extent of the ablation.
If the device is being used in a laparoscopic procedure, care must be taken to avoid
a gas embolism.
If the device is being used in a laparoscopic procedure, activation of the device when
not in contact with target tissue may cause capacitive coupling.
If the RF Generator shuts down for a ThermoPad over temperature error or if the
generator is shut off or rebooted for any reason, the ThermoPads must be replaced
and moved at least 4cm from their original location.
3.4 Warnings and Precautions Specific to the RITA
System
Electric shock hazard. Do not remove the cover of the RF Generator. Refer all
service to AngioDynamics, Inc. There are no user-serviceable parts inside the RF
Generator. Warranty will be voided if the unit is opened and/or the warranty
seal is broken.
Only use metal introducers that have insulation. RF energy can be transmitted from
the electrode through the un-insulated metal introducer to the patient causing
inadvertent burns.
Low power output or failure of the electrosurgical equipment to function correctly at
normal settings may indicate faulty application of the Dispersive Electrode or failure
of an electrical lead. Do not increase power output before checking for obvious
defects or misapplication. For monopolar surgery, effective contact between the
patient and the Dispersive Electrode must be verified whenever the patient is
repositioned.
Although accessories may have similar connector types, potentially hazardous
conditions may exist when inappropriate accessories are combined, and may cause
harm to the patient, user, or Generator. Therefore, it is essential that ONLY RITA
Electrosurgical Devices and accessories are used with RITA
RF Generators, and
there is proper connection between them.
Reusable accessory cables should be periodically tested for function and safety in
accordance with the cable’s instructions.
The use and proper placement of a Dispersive Electrode is a key element in the safe
and effective use of monopolar electrosurgery, particularly in the prevention of burns.
Page view 6
1 2 3 4 5 6 7 8 9 10 11 12 ... 29 30

Comments to this Manuals

No comments