AngioDynamics, Inc. Model 1500X User’s Guide and Service Manual
160-103990 Rev 03 Page 25
SYSTEM FAILURE 15: RF PULSE: HI
VOLTAGE
TURN POWER OFF
Turn Power off. Contact Customer Service.
SYSTEM FAILURE 16: RF PULSE: LO
VOLTAGE
TURN POWER OFF
Turn Power off. Contact Customer Service.
10.1.2 Troubleshooting Messages During Use
The following messages may appear on the LCD if an error occurs during use of the RF
Generator:
LCD Display Solution
“NO DEVICE PRESENT
Ensure that Device is properly connected to the
Device port. If problem persists, use alternate
Device cable or alternate Device. If problem
persists, contact Customer Service.
“BAD DEVICE OR CONNECTION / TC1----”,
“BAD DEVICE OR CONNECTION / TC-2---”,
“BAD DEVICE OR CONNECTION / TC--3--”,
“BAD DEVICE OR CONNECTION / TC----4”,
“BAD DEVICE OR CONNECTION / TC----5”,…
“BAD DEVICE OR CONNECTION / TC12345”
Check connection on device or replace device cable
or device.
“DEVICE NOT RECOGNIZED
BAD DEVICE OR CONNECTION / TC-----”
Ensure that Device is properly connected to the
Device port. If problem persists, use alternate
Device cable or alternate Device. If problem
persists, contact Customer Service.
“BAD DEVICE OR CONNECTION / NO ID, NO
TCs
1) Check pad and cable connections by un-plugging
and re-plugging in the device and/or Main Cable and
Dispersive Pad and/or Pad Adapter Cable, then
press “RF ON/OFF” to start.
2) If RF does not come back on as in one above, try
replacing the device and/or Main Cable, then press
RF ON/OFF to start.
“TARGET TEMP NOT REACHED”
Target temperature was not achieved (See Device
IFU).
“RE-HEATING TO TARGET TEMP”
Temperatures have decreased below target
temperature. (See Device IFU).
“RF WAS TURNED OFF
The user manually turned off the RF energy. If RF
energy delivery is desired, press the RF ON/OFF
button or press the foot pedal.
“POOR CONNECTION”
Needle tissue contact is out of the allowable range;
RF energy cannot be delivered. (See Device IFU). If
problem persists, contact Customer Service.
“RF SHORT
Needle tissue contact is out of the allowable range;
RF energy cannot be delivered. Check system set-
up. If problem persists, contact Customer Service.