AngioDynamics, Inc. Model 1500X User’s Guide and Service Manual
160-103990 Rev 03 Page 8
power can be increased and decreased. (Note: Set power as low as possible
for intended purposes.)
6. “DELIVERED POWER (W)” Display. Displays the actual power being delivered
in whole units of Watts.
7. “TIMER (min)” Display. Displays the time to a resolution of 0.1 minutes. This
display, prior to the start of the RF energy, shows the time set for RF energy
delivery at the target temperature. Once the RF energy starts, the display shows
the remaining time of RF energy delivery at the set temperature. If the Device
temperature is not maintained at the target temperature, the timer stops. The
timer resumes counting down once the target temperature is reached again.
During the cool down cycle this display counts up 0.5 minute to indicate the
duration of this cycle. Timer display is blank during track ablation.
7. “TIMER (min)” Buttons. Pressing the arrow buttons, sets the time the RF energy
will be on while at the target temperature. When this time has counted down to
zero, the RF energy delivery ceases. The timer can be increased and
decreased.
8. “RF TIME (min)” Display. Displays the total time the RF power has been on to a
resolution of 0.1 minutes. This display resets to zero at the onset of a new RF
energy delivery cycle. A RF energy delivery cycle is considered to be complete if
in the previous cycle the Timer reached zero or if the mode is changed.
9. “EFFICIENCY” Display. Displays the real time efficiency value on the tissue.
The display shows efficiency range of 0 to 10 (0 being lowest and 10 being the
highest). If the efficiency value is 0, the RF will not be activated. When RF energy
is being delivered, the desirable range of efficiency is 6 to 10, except for Track
Ablation mode where the acceptable efficiency range is 1 to 10.
10. “DEVICE TEMPERATURES C)” Display. Displays the temperature readings
of the device thermocouples in whole units of °C. For temperatures out of range
the display will either read “LO” or “HI”. Also, if using a Device with fewer than
five thermocouples, the locations with no thermocouple will display “OP”. If the
Device is not connected, the temperature displays remain blank.
10. “DEVICE TEMPERATURES C)” Buttons and LED’s. Each Device
temperature display has an accompanying button with a number on it and a
green LED. Pressing the button switches the LED ON and OFF. When the LED
is ON, the reading of that temperature sensor is used in the temperature control
algorithm. If the LED is OFF, the displayed value is not used in the calculation of
the average device temperature. The last display cannot be deselected in ATC
mode. The displays that indicate “OP” prior to the activation of RF energy are
excluded from the temperature algorithm in ATC mode and their LED’s are OFF.
11. “AUXILIARY C) TEMPERATURES” Display. Functions similar to the Device
Temperature Buttons and LED’s.
12. Liquid Crystal Display (LCD) Display. Displays the RF Generator’s current
status and operating information.