AngioDynamics, Inc. Model 1500X User’s Guide and Service Manual
160-103990 Rev 03 Page 19
9. SPECIFICATIONS
9.1 RF Generator Specifications
Constant Power, Constant Temperature, or Track Ablation
250 W into 40 – 60 Ω
. Power drops outside this impedance range.
(Refer to Figures 13.1 and 13.2.) Power accuracy (25 – 150 Ω) is ±20%
or 2 W, whichever is greater.
Range of 0 – 10.
Accuracy
±3°C from 15 - 125°C
±5°C below 15°C and above 125°C
135 VRMS into 90 Ω
460 kHz, ±5%
Accessories Rated Voltage
The accessories are rated for the maximum peak output voltage as
indicated in Figure 3.
100 - 240 V, 50 - 60 Hz, Auto-Switching Power Supply
6 Amp.
600 VA
Two 6.3 Amp 250 volt fuses (in the Power Entry Module on rear panel)
14.75" x 17.0” x 5.25” (width x depth x height)
(37.5 cm x 43 cm x 13.5 cm)
23 lbs. (10 kg)
Power on/off, RF on/off, Control Mode Set, Target Temperature Set,
Power Set, Time of Energy Delivery Set, Activate/Deactivate
Temperature sensors, Track Ablation On/Off, Purge On/Off
Target Temperature, Power Setting, Timer, Delivered Power, Time that
RF is delivered, Efficiency, Temperatures for all Device and Auxiliary
Thermocouples, Message Center.
Foot pedal Port, Main Cable Port (14-pin polarized connector to
electrodes with thermocouples), Dispersive Electrode Ports, RS232 Data
Ports, AUX Port (6-pin connector), Power Entry Module, Software
Module.
Class 1, Defibrillator Proof –
Type BF for ordinary, continuous operation.
This equipment is not suitable for use in the presence of a flammable
anesthetic mixture with air, Oxygen, Nitrous Oxide.
DB-9 connectors for RITA
base and Intelliflow pump
Temperatures: -20°C to +50°C
Humidity: 20 - 85%, non-condensing