AngioDynamics, Inc. Model 1500X User’s Guide and Service Manual
160-103990 Rev 03 Page 9
Connections/Ports:
13. “FOOT PEDAL” Port. Port for connecting a pneumatic foot pedal (for activating
and deactivating RF energy delivery). The foot pedal functions like RF ON/OFF
switch. The foot pedal may also be used to start and stop the purge operation
when the RF Generator is used with an infusion pump.
14. “RETURN” Dispersive Electrode Port. Port for connecting the Dispersive
(Return) Electrode(s) from the patient to the RF Generator.
15. “DEVICE” Port. Port for connecting the Device to the RF Generator via the
Main Cable. The port is keyed to ensure proper connection.
16. “AUX” Port. The port is keyed (different from the Device port) to ensure proper
connection.
4.2 Rear Panel
Switches and Connections:
1. RS-232. Serial ports for connecting to external devices for data communication.
Any device connected to RS232 data ports must comply with the requirements of
IEC 60601-1.
Port A: Port for connecting to a personal computer for use with RITA
-Base
Data Collection Software.
Port B: Port for controlling a RITA
specific infusion pump for use with the RF
Generator.
Port C: Port not defined/for future use. Not for user operation.
2. Software Module Access. The main software for the operation of the generator
resides in a Software Module. The RF Generator comes with this software
module already installed. If replacement of the Software Module is required
(e.g., repair, upgrade, etc.), this Module can be accessed by removing the
Software Module Access cover. This should only be done by qualified, resident
bioengineers and technicians. Refer to Section 7.1.