AngioDynamics, Inc. Model 1500X User’s Guide and Service Manual
160-103990 Rev 03 Page 15
In the track ablation mode “TRACK ABLATION” message will appear on the LCD. Note:
The efficiency value (0-10) is not important to the efficiency of ablating the track.
Set Power if desired power level is different from the default. Set power as low as
possible for intended purpose.
When ready to start, turn the RF energy on by pressing the RF ON/OFF” button or
depressing the foot pedal.
Watch the temperatures. As the highest temperature reaches 80°C, the power will
automatically decrease to maintain the highest temperature reading at 80°C. Pull the
device back 1 cm. Continue to pull back 1 cm, using the centimeter marks on the
Device, each time 80°C is reached.)
Turn off RF power when done.
WARNING: Ensure that RF power is turned off after Track Ablation. Failure to do so
could result in unintended tissue ablation.
6.9 Bipolar Mode:
This mode is used when using a resectioning device such as Habib
4X. In this mode ground
return electrodes are not used and a special plug adapter brings RF and RF return to the
Device. An example of a special plug adapter is shown below.
7. CLINICAL STUDIES
To review clinical studies please go to:
http://www.angiodynamics.com/pages/physicians/us_trials.asp (if in the U.S.)
http://www.angiodynamics.com/pages/physicians/int_trials.asp (if outside the U.S.)